• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® SPINAL ANESTHESIA TRAY; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® SPINAL ANESTHESIA TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number A2873
Device Problem Missing Value Reason (3192)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 07/03/2017
Event Type  malfunction  
Manufacturer Narrative
Visual evaluation of marcaine-bupivacaine ampules from ten trays, returned by the customer, found no visible non-conformities.The drug liquid was found to be clear (colorless) with no particulates visible inside the vial.Review of device history records and incoming records found no discrepancies or anomalies relevant to the complaint.Assay testing, performed by the supplier prior to release of bupivacaine-marcaine lot # 700503a, was within specification.Review of the sterilization certificate for the tray lot number 3387507 (po # (b)(4)) revealed that the product was processed according to validated specification requirements and parameters.Smith medical periodically monitors supplied drug components stability post ethylene oxide sterilization throughout their expiry.Our records show that the supplied drugs components continue to meet potency specifications.Complaint history review does not indicate any upward trends for the drug lot number, reported in this complaint.The anesthetics are supplied items and the complaint notification was forwarded to the supplier.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Event Description
It was reported that during use of a portex® spinal anesthesia tray bupivacaine was not effective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SPINAL ANESTHESIA TRAY
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL NORTH AMERICA
10 bowman drive
keene NH 03431
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key7508268
MDR Text Key108158643
Report Number3012307300-2018-10297
Device Sequence Number1
Product Code CAZ
UDI-Device Identifier00351688055739
UDI-Public00351688055739
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2019
Device Catalogue NumberA2873
Device Lot Number3387507
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/06/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-