• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 5-EX SCANNER; TRANSDUCER, ULTRASONIC, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 5-EX SCANNER; TRANSDUCER, ULTRASONIC, DIAGNOSTIC Back to Search Results
Catalog Number 9763000
Device Problem Computer Operating System Problem (2898)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.Device not returned.
 
Event Description
It was reported that the power may be turned off during use.Additional information supplied on (b)(6) 2018: "even if the hospital does not operate anything, this site light will be powered down.It seems that the screen becomes dark and the power supply is dropped in a state with charging.".
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: during evaluation, the reported issue of the system powering-off or becoming dark during use could not be reproduced.The scanner powers on, boots up normally and has a good image for the current version of software.The review of the dhr, sub-assembly dhr, sub-assembly manufacturing, component records, manufacturing process changes, mrr/mrb, and internal rejects showed that the reported issue is not likely related to a manufacturing issue.No other issues with the equipment were found during evaluation.The sr5 was serviced, qc tested, and returned to the customer.
 
Event Description
It was reported that the power may be turned off during use.04/25/2018 additional information: "even if the hospital does not operate anything, this site light will be powered down.It seems that the screen becomes dark and the power supply is dropped in a state with charging.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 5-EX SCANNER
Type of Device
TRANSDUCER, ULTRASONIC, DIAGNOSTIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
sydney freckleton
605 n. 5600 w.
salt lake city, UT 84116
8015225992
MDR Report Key7508667
MDR Text Key108139971
Report Number3006260740-2018-00958
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741091032
UDI-Public(01)00801741091032
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9763000
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Event Location Hospital
Date Manufacturer Received10/24/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-