• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-27
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, there was a reported missing sensor wire.The sensor was inserted into the abdomen on (b)(6) 2018.The device has been received for investigation.A follow-up will be submitted upon completion of device investigation.Confirmation of the problem and probable cause could not be determined.No additional event or patient information is available.
 
Manufacturer Narrative
Com-(b)(4).
 
Event Description
A sensor was returned for evaluation.A visual inspection was performed and found that the sensor wire is missing from the sensor pod and seal carrier.The reported event of a missing or detached wire was confirmed.A probable cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of Device
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key7511294
MDR Text Key108318855
Report Number3004753838-2018-051397
Device Sequence Number1
Product Code MDS
UDI-Device Identifier00386270000019
UDI-Public00386270000019
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/26/2019
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Device Lot Number5234618
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/14/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age70 YR
Patient Weight95
-
-