• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. STRATAFIX SPIRAL PDS+; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. STRATAFIX SPIRAL PDS+; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Catalog Number SXPP1B450
Device Problem Failure To Adhere Or Bond (1031)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that a patient underwent hysterectomy procedure on (b)(6) 2018 and barbed suture was used.The surgeon opined that the barbs are more shallow and do not hold the tissue as well as similar devices.There were no patient consequences reported.
 
Manufacturer Narrative
(b)(4).No product was returned for evaluation to determine the assignable cause of the performance barb non-engagement in tissue ; therefore, the reported failure could not be evaluated.It could not be determined what may have caused the reported incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX SPIRAL PDS+
Type of Device
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.
road 183, km. 8.3
san lorenzo PR
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville, NJ 08876-0151
9082182792
MDR Report Key7515150
MDR Text Key108288469
Report Number2210968-2018-72801
Device Sequence Number1
Product Code NEW
UDI-Device Identifier10705031236219
UDI-Public10705031236219
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue NumberSXPP1B450
Device Lot NumberLLH578
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/08/2018
Date Device Manufactured10/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-