Model Number 03.501.080
Device Problem Sticking (1597)
Patient Problem No Patient Involvement (2645)
Event Date 04/25/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.A device history records review has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the sternal zipfix application instrument is sticking.The issue was observed during pre-surgery with no patient involvement.This report is for an application instrument for sternal zipfix.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Part 03.501.080, lot l254389: manufacturing location: haegendorf.Release to warehouse date: january 26, 2017.No non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.A service & repair evaluation was completed: the customer reported the device was sticking.The repair technician reported the retaining nut and handle screw were loose, and the trigger was sticky.Lube/oil/clean is the reason for repair.The cause of the issue is unknown.No parts were replaced.The item was repaired per the inspection sheet, passed synthes final inspection and returned to the customer.The evaluation was confirmed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7516054
MDR Text Key108359893
Report Number2939274-2018-52189
Device Sequence Number1
Product Code JDQ
UDI-Device Identifier10887587010847
UDI-Public(01)10887587010847
Combination Product (y/n)N
PMA/PMN Number
K110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.501.080
Device Catalogue Number03.501.080
Device Lot NumberL254389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2018
Date Manufacturer Received06/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-