Catalog Number BI70002000
Device Problem Image Reversal (1358)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/18/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative noted they were not able to replicate the issue.The imaging system then passed the system checkout and was found to be fully functional.No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding an imaging device being used for a spinal fusion procedure.It was reported that the radiologic technologist picked the correct orientation but when it came up on the mobile viewing station, it was backwards.When they sent the image to the navigation system, it was the correct orientation.There was no reported impact on patient outcome.There was no reported delay to the procedure due to this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
sarah olhoft
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7516492
MDR Text Key108362078
Report Number1723170-2018-02060
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Catalogue NumberBI70002000
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/05/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age42 YR
Patient Weight79
-
-