Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2018
Event Type  malfunction  
Manufacturer Narrative
The reported event was reproduced during functional testing of the autopulse platform (sn (b)(4)).Review of the archive data confirmed the occurrence of multiple user advisory (ua) 16 (timeout moving to take-up position) error messages on (b)(6) 2018 and not on the reported event date of (b)(6) 2018.This confirms the customer reported event.The platform was functionally tested and during initial take up, displayed a (ua) 16 (timeout moving to take-up position) error message which was attributed to a seized brake gap.The root cause of the ua 16 error message is due to this observed issue.After the brake gap was unseized, the platform was further functionally tested including the load characterization check and passed all final specification.Historical complaints were reviewed for service information related to the reported complaint and no previous history of complaint was reported for the autopulse platform with serial number (b)(4).
 
Event Description
During a product demonstration, the autopulse platform displayed a user advisory (ua) 16 (timeout moving to take-up position) error message.The user unsuccessfully attempted to clear the ua 16 error message by reinstalling the lifeband.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7524069
MDR Text Key108618262
Report Number3010617000-2018-00451
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-