(b)(4).The sample was returned and sent on the manufacturer (pegasus research corp).Pegasus reports that upon receipt of unit it was found that the heater was functional.The heater was tested for electrical safety and temperature output, and found to be within specifications.The unit was left on, with and without open case, for eight consecutive hours on each case, and the unit did not present variations in temperature out of the ones in specifications.The approximate age of the heater is 277 days.Based on the investigation performed, the complaint could not be confirmed.There were no issues found with the returned sample.
|