COVIDIEN 840 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 840
Device Problem Device Inoperable (1663)
Patient Problem No Code Available (3191)
Event Date 04/21/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation summary: the field service engineer (fse) evaluated the ventilator and found a code in the ventilator memory logs but could not duplicate the reported event.The ventilator passed all testing and was returned to clinical use.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that, while in use on a patient, 840 ventilator did not switch to battery when there was a power outage.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
Manufacturer Contact
kelly adams
2101 faraday avenue
carlsbad, CA 92008
7606035046
MDR Report Key7530786
MDR Text Key108864010
Report Number8020893-2018-00234
Device Sequence Number1
Product Code CBK
UDI-Device Identifier10884521181045
UDI-Public10884521181045
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151252
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number10049998
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2018
Initial Date Manufacturer Received 04/30/2018
Initial Date FDA Received05/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-