• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3572-1
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 04/24/2018
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported that the torque output of a torque limiting handle was found outside of the adequate performance range.This was detected during a preliminary calibration check by zimmer biomet spine personnel.There are no specific surgical procedures associated with this handle.
 
Manufacturer Narrative
The returned device was evaluated.The torque output was confirmed to be outside of the range.There were no signs of abnormal wear found.Handles found out of calibration typically exhibit spring relaxation, wear of critical torque components, and/or break down of the internal lubrication from use over time.A review of the manufacturing records did not identify any issues which may have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR HANDLE, T RATCHETING TLR 90 IN-LBS
Type of Device
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7530851
MDR Text Key108860483
Report Number3012447612-2018-00425
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK132884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number3572-1
Device Lot Number76LA-024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received09/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-