Catalog Number 8065750833 |
Device Problems
Occlusion Within Device (1423); Device Displays Incorrect Message (2591)
|
Patient Problems
Burn, Thermal (2530); No Code Available (3191)
|
Event Date 04/26/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A customer reported that during a cataract procedure, the phaco tip occluded and a patient experienced a wound burn, in the left eye.The surgeon was removing the tip from the eye to flush and got to close to the incision while still in position 3 on the foot pedal.Sutures were used to close the wound.Postoperatively, the patient is reported as doing "fine." the phaco tip was not saved for an evaluation.Additional information was received from the customer indicating the occlusion tone was a continuous ding alarm and the machine noted an occlusion.A contact lens was placed in the patient's eye four days postop due to small aqueous leak.The patients outcome is reported as very good.
|
|
Manufacturer Narrative
|
No phaco tip sample was returned for evaluation for the report of a thermal burn due to the tip becoming occluded; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a lot history review could not be conducted.Because a sample was not returned from the customer and no lot information was provided for a lot record review, the root cause for customer complaint issue cannot be determined.Phaco tips are 100% visually inspected by trained operators during the manufacturing process.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|