• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750833
Device Problems Occlusion Within Device (1423); Device Displays Incorrect Message (2591)
Patient Problems Burn, Thermal (2530); No Code Available (3191)
Event Date 04/26/2018
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that during a cataract procedure, the phaco tip occluded and a patient experienced a wound burn, in the left eye.The surgeon was removing the tip from the eye to flush and got to close to the incision while still in position 3 on the foot pedal.Sutures were used to close the wound.Postoperatively, the patient is reported as doing "fine." the phaco tip was not saved for an evaluation.Additional information was received from the customer indicating the occlusion tone was a continuous ding alarm and the machine noted an occlusion.A contact lens was placed in the patient's eye four days postop due to small aqueous leak.The patients outcome is reported as very good.
 
Manufacturer Narrative
No phaco tip sample was returned for evaluation for the report of a thermal burn due to the tip becoming occluded; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a lot history review could not be conducted.Because a sample was not returned from the customer and no lot information was provided for a lot record review, the root cause for customer complaint issue cannot be determined.Phaco tips are 100% visually inspected by trained operators during the manufacturing process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7531050
MDR Text Key108856142
Report Number2028159-2018-01045
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number380657508334
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/15/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OZIL TORSIONAL PHACO HP; TS TIP,45KT,MINI FL AB 0.9MM
Patient Outcome(s) Other; Required Intervention;
Patient Age72 YR
-
-