MEDTRONIC IRELAND MICRA; INTRODUCER, CATHETER Back to Search Results
Model Number MI2355A
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, aspiration during use of the introducer sheath continuously attracted air.The introducer sheath was replaced with a different product.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the full introducer was returned and analyzed.Analysis indicated that the lumen of the catheter was kinked/buckled.Blood obstruction of the delivery system lumen was observed.Visual analysis of the lead indicated damage during use.The analyst noted that the flush tube kinked could be contributed to the complaint of aspiration attracted air during the procedure.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRA
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7531147
MDR Text Key108857059
Report Number9612164-2018-01175
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K132030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/11/2019
Device Model NumberMI2355A
Device Catalogue NumberMI2355A
Device Lot Number00126766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-