Investigation evaluation: a product evaluation was performed only by the pictures provided in response to this report because the product said to be involved was not provided to cook for evaluation.The photo provided is said to be for either mdr 1037905-2018-00214 (related mdr) or 1037905-2018-00215.Based on the photo provided, we can confirm that the forceps cups are in an open position and the open cups appear not to be symmetrical; however, we cannot confirm whether or not the cups will open and close as intended.Without substantial evidence to contradict the complaint, it is considered confirmed based solely on statements describing the event.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use state the following in regards to product inspection: "beginning at the handle and moving toward the cups, uncoil the forceps making sure not to stretch the cable.Open and close the cups to verify smooth handle operation and appropriate cup action.Become familiar with the amount of handle movement required to operate the cups.If any irregularities are noted, do not use.Note: exercising the handle while the forcep is coiled may result in damage to the performance characteristics of the forceps." the instructions for use state, "with cups closed, insert forceps into accessory channel.Note: keep end of forceps extending from accessory channel straight at all times.Allowing forceps to hang from accessory channel may cause damage to forceps." prior to distribution, all captura serrated large forceps- spike are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
During biopsy procedures, the physician used twelve (12) captura serrated large forceps - spike.The biopsy forceps did not close properly.When the physician tried to open the forceps before taking the biopsy, the mechanism was found to be broke.They used another one without any problem.
|