Model Number 840 |
Device Problem
Device Inoperable (1663)
|
Patient Problem
No Code Available (3191)
|
Event Date 04/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device evaluation the field service engineer (fse) evaluated the ventilator and replaced the proportional solenoid valve, inspiratory flow sensor and exhalation flow sensor.The ventilator passed all testing and was placed back in clinical use.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that, while in use on a patient, a 840 ventilator generated an error code and stopped ventilating.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.
|
|
Manufacturer Narrative
|
Device evaluation: the proportional solenoid (psol) was returned for failure investigation.The psol was installed into a failure investigation test ventilator for analysis.The unit powered up normally and no errors were recorded in the diagnostic logs.The f.I.Ventilator was put into ventilation mode for a minimum of 24 hrs using default settings as per 840 service manual.No failures occurred.The psol was functionally tested and failed testing due to the high flow unable to be adjusted.The psol was removed and the coil cover, shunt, base and locknut were replaced.The unit was once again functionally tested but failed the same symptoms.The spring pack was replaced and the psol retested.The unit passed all tests.A visual inspection of the removed spring pack assembly revealed that the armature body was not sitting correctly on the assembly.Compared to other assemblies in the manufacturing area this armature body appeared to be sitting higher on the assembly.It is not possible to breakdown the unit because during the assembly it is pneumatically p ressed to a maximum force of between 1660 and 1780lb of pressure.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional codes added to evaluation codes result and conclusion.Device evaluation summary: the inspiratory flow sensor was returned for failure investigation.A visual inspection of the returned part showed no anomalies.The unit was installed into a failure investigation (fi) test ventilator for analysis.There was no errors observed.The fi ventilator was put into ventilation mode for a minimum of 24 hrs using default settings as per 840 service manual.No failures occurred.The exhalation flow sensor was returned for failure investigation.A visual inspection of the returned part showed no anomalies.The unit was installed into a fi for analysis.There was no errors observed.The fi ventilator was put into ventilation mode for a minimum of 24 hrs using default settings as per 840 service manual.No failures occurred.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|