Model Number MSERIES |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device displayed a "pacer fault 117" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll france.The malfunction was observed during incoming testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.Analysis for reports of this type has not identified an increase in trend.The device was returned to the customer labeled not for clinical use.Analysis of reports of this type has not identified an increase in trend.
|
|