BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The exact age of the patient is unknown.However, it was reported the patient was over 18 years.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold¿ lite with capio slim was used during a colpopexy and colporrhaphy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle of the capio was ¿released¿ outside the patient.The procedure was completed with another uphold¿ lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
A visual examination of the returned uphold lite with capio slim revealed that the darts were not detached from either of the two sutures.Both darts were caught on the cage of the capio slim when actuated.However, the carrier exited the capio slim head slightly bent and then righted itself to the cage which would cause the carrier to stay on a misaligned trajectory during use while going through bodily tissue causing it to miss the cage.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.Based on the investigation finding that the neither dart had detached from its respective suture, this is no longer considered a reportable event.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a colpopexy and colporrhaphy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle of the capio was ¿released¿ outside the patient.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|