BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/24/2018
Event Type  malfunction  
Manufacturer Narrative
The exact age of the patient is unknown.However, it was reported the patient was over 18 years.(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold¿ lite with capio slim was used during a colpopexy and colporrhaphy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle of the capio was ¿released¿ outside the patient.The procedure was completed with another uphold¿ lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual examination of the returned uphold lite with capio slim revealed that the darts were not detached from either of the two sutures.Both darts were caught on the cage of the capio slim when actuated.However, the carrier exited the capio slim head slightly bent and then righted itself to the cage which would cause the carrier to stay on a misaligned trajectory during use while going through bodily tissue causing it to miss the cage.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.Based on the investigation finding that the neither dart had detached from its respective suture, this is no longer considered a reportable event.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during a colpopexy and colporrhaphy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the needle of the capio was ¿released¿ outside the patient.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7534279
MDR Text Key108992324
Report Number3005099803-2018-01488
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2019
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000047878
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/25/2018
Initial Date FDA Received05/22/2018
Supplement Dates Manufacturer Received08/17/2018
Supplement Dates FDA Received09/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-