• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA APOLLO; ANESTHESIA UNITS Back to Search Results
Catalog Number 8606500
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/19/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event Description
It was reported there was a mixer and ventilator failure during a case.There no injury reported.
 
Manufacturer Narrative
The electronic log file was evaluated.A sporadic malfunction of a cpu board which controls the device-internal communication between user interface and gas mixer could be clearly identified as the root cause.Dräger knows a certain number of similar cases where a sporadic breakdown of device-internal communication occurred.This leads to a situation where ventilator and gas mixer enter a fail state simultaneously; the device is designed to switch to monitoring mode then and to alert the user to this condition by means of a corresponding alarm.Manual ventilation with emergency oxygen dosage remains possible, this includes application of anesthetic gas as well.The monitoring functionalities remain unaffected.The procedure how to establish the emergency gas supply is laid down in the ifu.In all comparable cases it was only possible to narrow down the root cause to the respective pcb cpu - the exact failure mechanism however could not be determined by in-depth analysis.The fact that the identical type of board is used in the workstation twice and does not exhibit malfunctions in the second application periphery makes a general design issue rather unlikely.A reasonable explanation would be electrostatic discharge of the user during interaction with the device or any other kind of electromagnetic disturbance that exceeds the immunity barriers of the device.The primus was developed in compliance to the requirements of iec 60601-1-2.The number of similar cases is within the expected range of the respective risk assessment and thus accepted.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APOLLO
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key7536356
MDR Text Key109256609
Report Number9611500-2018-00147
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
PMA/PMN Number
K042607
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/01/2000
Device Catalogue Number8606500
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/23/2018
Initial Date FDA Received05/23/2018
Supplement Dates Manufacturer Received05/04/2018
Supplement Dates FDA Received06/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-