• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM D; HIP PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM D; HIP PROSTHESIS Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.(b)(6).Concomitant products: biomet g7 finned bm 3 hole shell 50d, item#: 110017122, lot#: 6223319.
 
Event Description
It was reported that during hip arthroplasty, a liner would not fix into the shell properly.A backup liner was used to finish the surgery.No known adverse event was reported.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Updated: complaint sample was evaluated and the reported event was not confirmed.Upon visual inspection there is a circular indentation on the outside radius.There was no further damage to either outside or the inside of the liner.The device was fully inspected and confirming to print specifications.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM D
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7536368
MDR Text Key109003812
Report Number0001825034-2018-03320
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/06/2023
Device Model NumberN/A
Device Catalogue Number010000848
Device Lot Number6207321
Other Device ID Number(01) 0 0880304 52634 1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2018
Date Manufacturer Received09/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-