Model Number N/A |
Device Problems
Difficult to Insert (1316); Difficult To Position (1467)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.(b)(6).Concomitant products: biomet g7 finned bm 3 hole shell 50d, item#: 110017122, lot#: 6223319.
|
|
Event Description
|
It was reported that during hip arthroplasty, a liner would not fix into the shell properly.A backup liner was used to finish the surgery.No known adverse event was reported.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Updated: complaint sample was evaluated and the reported event was not confirmed.Upon visual inspection there is a circular indentation on the outside radius.There was no further damage to either outside or the inside of the liner.The device was fully inspected and confirming to print specifications.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|