• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 195240
Device Problems Failure To Adhere Or Bond (1031); Out-Of-Box Failure (2311)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/30/2018
Event Type  malfunction  
Manufacturer Narrative
The reported complaint was not verifiable.The defective pads have been disposed of by the user facility.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the level sensor pads were not sticking.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review: during a cpb procedure on (b)(6) 2018, the level sensor pad would not stick appropriately on their sorin/livanova inspire oxygenator, rounded integrated reservoir.The team went through four sensors before and during cpb to get one that adhered appropriately to sufficiently hold the level sensing cable, and have the system work to specifications.There was not a concern about either the cable or the level module.It is unsure if the team is waiting the instruction for use (ifu) described five minutes before they load their cable, and if they are placing the sensor pad on the labeled/sticker area on the rounded reservoir.The account is a long time user of the system and we have discussed the ifu on the steps top correctly engage the level sensing system.The incident did not delay the surgical procedure, and there was no blood loss or harm associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7536527
MDR Text Key109044376
Report Number1828100-2018-00273
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163531
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/23/2018
Device Model Number195240
Device Catalogue Number195240
Device Lot Number822508
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/01/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SORIN/LIVENOVA INSPIRE OXYGENATOR
-
-