• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTRAVIA CONTAINER EMPTY; CONTAINER, I.V. Back to Search Results
Catalog Number 2B8011
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/05/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that it was difficult to remove an intravia container empty from tubing, and the bag broke during removal.The reporter stated forceps were used to remove the part of the bag stuck to the tubing.It was not specified when in the process this event occurred.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed and it was observed that the membrane tube assembly was detached from the bag.A functional test was performed and the bag was submerged underwater and no leakage was observed.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA CONTAINER EMPTY
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina, san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7541733
MDR Text Key109282055
Report Number1416980-2018-03148
Device Sequence Number1
Product Code KPE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/05/2023
Device Catalogue Number2B8011
Device Lot NumberDR18A03085
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-