• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERMO FISHER SCIENTIFIC OY THERMO SCIENTIFIC¿ TCAUTOMATION¿; LABORATORY AUTOMATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERMO FISHER SCIENTIFIC OY THERMO SCIENTIFIC¿ TCAUTOMATION¿; LABORATORY AUTOMATION SYSTEM Back to Search Results
Catalog Number 952040
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913); Patient Data Problem (3197)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/18/2018
Event Type  malfunction  
Manufacturer Narrative
Based on the investigation results the it was determined that after the tube for sample a was placed in the centrifuge module load rack, the operator stopped the centrifuge module but did not remove all the samples as required per the instructions in the user manual.The operator then started the centrifuge module by releasing the module lock but did not remove all the samples prior to releasing the module lock as required per the instructions in the user manual.The tcautomation software performed as designed by correctly identifying the sample identification (sid) mismatch by posting a cross check error message.The results of the investigation did not identify any failures with the system.
 
Event Description
A customer reported that test results from a patient sample processed on the thermo scientific¿ tcautomation¿ laboratory automation system were mis-associated with the sample identification number (sid) of a different patient.Mis-associated patient results may lead to inappropriate physician action.The mis-associated patient test result was reported from the laboratory.There is no allegation of patient harm as a result of this event.The tcautomation software correctly identified the sid mismatch by posting a cross check failure message, as designed.This report corresponds to thermo fisher scientific (b)(4) complaint number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMO SCIENTIFIC¿ TCAUTOMATION¿
Type of Device
LABORATORY AUTOMATION SYSTEM
Manufacturer (Section D)
THERMO FISHER SCIENTIFIC OY
ratastie 2
vantaa, 01620
FI  01620
Manufacturer Contact
annika astola
ratastie 2
vantaa, 01620
FI   01620
MDR Report Key7543821
MDR Text Key109886563
Report Number9610942-2018-00001
Device Sequence Number1
Product Code LXG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Type of Report Initial
Report Date 05/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number952040
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-