Model Number 16-02-85 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/26/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information was not provided.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.Device not returned.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t lost power during a procedure.The customer reported that the device switches off after 30 seconds of run time.There was no report of patient injury.
|
|
Manufacturer Narrative
|
A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and replaced the power switch, power supply board and the patient bridge to resolve the malfunction.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
The device was returned for further investigation.During the evaluation the reported issue could not be confirmed or reproduced.The device worked according the specification.As the issue could not be reproduced or confirmed, a root cause was not identified.
|
|
Event Description
|
See initial.
|
|