The customer strips were returned for investigation.The returned test strips were measured with the current test strip master lot.For this purpose two human blood samples from marcumar donors and internal reference meters were used.Testing results: donor 1: reference meter with masterlot strips: 2.4 inr, reference meter with customer strips: 2.4 inr.Donor 2: reference meter with masterlot strips: 7.0 inr, reference meter with customer strips: >8.0 inr.The maximum difference between measurements with the same blood sample was 14 %.Returned customer material and retention material comply with specification.
|