• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT HPS BPH FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT HPS BPH FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2090
Device Problems Thermal Decomposition of Device (1071); Material Perforation (2205); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/26/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis of the retuned device revealed that the fiber cap exhibits signs of melting which formed a hole from the surface to the bevel edge.The glass cap exhibits devitrification at the output area and detritus adhesion around output area.The heat shrink tubing exhibits minor scratch marks.Based on the device analysis, the potential for forward firing may exist.The probable cause of the failure is heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.An evaluation conclusion code of operational context was assigned to this investigation.
 
Event Description
Analysis of the returned device revealed that the fiber cap exhibits signs of melting which formed a hole from the surface to the bevel edge.Based on the device analysis, the potential for forward firing may exist.No patient clinical consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT HPS BPH FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7549848
MDR Text Key109484958
Report Number2937094-2018-00530
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953004709
UDI-Public00878953004709
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2019
Device Model Number0010-2090
Device Catalogue Number0010-2090
Device Lot Number21334128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-