The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The device was cut off at the insertion part of it.The loop was caught in the distal end other than the loop hanger which holds the loop.After removing the loop from the device, we checked the cutter which cuts the loop and the loop hanger of the device.As a result, there were no abnormalities.The manufacturing history record was reviewed, with no irregularities noted.This type of the event is most likely related to the operator's technique.Based on the past similar cases, it is assumed that the event occurred because the user pulled the operation handle of the device without positioning the loop vertically on the loop hanger.The instruction manual of the device has already warned as follows; *do not try to cut the loop that is not positioned on both edges of the loop hanger as plumb as possible for the blade.It may make cutting the loop impossible, or result in the loop getting caught in the distal end of the instrument, which could make it difficult or impossible to remove from the patient.In this case, use pliers to cut the insertion portion of the instrument where it extends from the biopsy valve of the endoscope.Remove the endoscope from the body, then reinsert the endoscope and cut the loop with a spare loop cutter.
|
During a colonoscopic polypectomy, the user tried to cut the loop inside the patient using the subject device.The user could not remove the device from the patient, since the loop was caught in the cutter of the device.The user used pliers to cut the insertion part of the device.The user removed the endoscope from the patient while leaving the device inside the patient.The user inserted the endoscope again, cut the loop with another device and removed the device from the patient.The intended procedure was completed.There was no patient injury reported.This is the report regarding the failure of cutting the loop.
|