Model Number X SERIES |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device displayed an "ecg disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical (b)(4) service department for evaluation and the customer's report was not duplicated under testing.Review of the device activity log does show occurrences of the reported error message.The multi-function cable gasket was replaced as a precaution.Analysis of reports of this type has not identified an increase in trend.
|
|