MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER WITH ENFIT¿; NG TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER WITH ENFIT¿; NG TUBE Back to Search Results
Catalog Number EN0046180
Device Problem Suction Problem (2170)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The reported issue could not be confirmed due to poor sample condition of the pictures.During the visual evaluation, a nasogastric tube with enfit connector assembly was noted.The received photos were reviewed and no kinks or obstructions were observed in the pvc tube or the enfit connector that could affect the flow rate or suction during the use.The sample could not be measured or tested since no physical sample was received.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "indications for use: bard® nasogastric sump tubes with enfit¿ connector are intended to be used for: decompression of stomach by suction or aspiration of gastric contents.Short-term administration of term tube feeding, lavage fluid and medications.Contraindications: patients with known tape or adhesive allergies.Warnings: use with caution in patients with a history of head trauma, facial trauma, esophageal diseases and patients with potential for vomiting.Do not force nasogastric tube during insertions; damage to the nasal passage and mucosa and bleeding may occur.Measure insertion length carefully- excessive insertion length of tube into the stomach may lead to coiling and/or formation of tube-knot.Lubricate the tube generously with water soluble lubricant prior to insertion.Do not use petroleum-based products as they may be harmful to the respiratory tract.Reflux of gastric contents into the blue vent lumen indicates that the suction lumen is obstructed or suction is too low.Routinely check for reflux in the blue vent lumen and clear as per applicable directions.Failure to clear the obstruction or clear prevent® filter may cause gas & fluid buildup in stomach, aspiration of gastric contents, aspiration pneumonia and other complications.Do not inject fluid through the prevent® filter as this may result in blockage and leakage of filter.Monitor patient for nasal erosion, sinusitis, esophagitis, esophagotracheal fistula, gastric erosion and pulmonary & oral infections.Statlock® nasogastric stabilization device: avoid contact with alcohol or acetone; both can weaken bonding of components and statlock® stabilization device pad adherence.Instructions for nasogastric tube insertion: explain the procedure to the patient.Carefully measure to find desired length of the tube using the nasogastric tube as a measurement aid.To determine the insertion length: measure the tube from the tip of the nose to the earlobe and from the earlobe to the tip of the xiphoid process.Mark the length of the tube to be passed with a small piece of tape.Check the patient¿s nostrils for patency; select the nostril with best patency.Lubricate the full length of tube to be inserted.Insert the tube through the nose aiming down and back.When the tube hits the pharynx, if patient is able, have him or her flex his/her head forward and swallow.Advance the tube as the patient swallows.If resistance is met, rotating the tube may facilitate placement.Continue to advance the tube until the marked position on the tube is reached.Do not advance beyond the marked length as coiling and or knotting of the tube in the stomach may occur.Confirm tube placement per hospital policy.The tube has a radiopaque stripe facilitating x-ray confirmation.If proper placement of tube within the stomach cannot be confirmed, remove the tube gently and start the procedure again.Secure with a securement device or tape per hospital protocol.If applicable, ensure the lopez valve® with enfit¿ connector is snugly inserted into suction lumen to prevent suction loss.Keep blue vent lumen above the level of the patient¿s stomach to prevent reflux of stomach fluids into the blue lumen.Do not clamp air vent port while suction is being applied.Recommended suction settings always use lowest suction setting that will effectively decompress the stomach.For intermittent suction via thermotic pump, use ¿high¿ (gomco, 120mm hg).For intermittent suction via central source, set at ¿low¿ (30-40mm hg).For continuous suction, set at ¿low¿ (30-40mm hg).Increase slowly until flow is observed as necessary.Instructions for prevent® anti-reflux filter with enfit¿ connector: firmly twist the white base of anti-reflux filter in blue air lumen vent of nasogastric tube.If gastric reflux in vent lumen is observed, clear the obstruction in the main lumen by following your hospital¿s standard protocol.Attach enfit¿ - compatible syringe to enfit¿ male fitting on anti-reflux filter and inject a minimum of 15cc of air to clear the blue air vent lumen of any gastric reflux.Do not inject fluid through filter.To cap nasogastric tube when tube is not connected to a suction source insert tethered cap that is attached to the barb in the suction lumen of nasogastric tube.Sterile unless package is opened or damaged, except for any individually packaged components within the pouch which are not labeled as sterile.These components are not terminally sterilized.Single use.Do not resterilize." the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the ng tube did not suction well.The complainant reported that the "inner moat" of the enfit connector limited the suction.The complainant tested the product without a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® NASOGASTRIC SUMP TUBE WITH PREVENT® ANTI-REFLUX FILTER WITH ENFIT¿
Type of Device
NG TUBE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7557526
MDR Text Key109766558
Report Number1018233-2018-01981
Device Sequence Number1
Product Code FEG
UDI-Device Identifier00801741113260
UDI-Public(01)00801741113260
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960176
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2019
Device Catalogue NumberEN0046180
Device Lot NumberNGAY2329
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/02/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-