ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40C
Device Problems No Device Output (1435); Over-Sensing (1438)
Patient Problem Bradycardia (1751)
Event Date 05/08/2018
Event Type  malfunction  
Event Description
It was reported that the cardiac resynchronization therapy defibrillator exhibited non-sustained oversensing due to myopotential oversensing and atrial cross talk.There was inhibition of pacing due to oversensing.Loss of capture was also noted.The device was reprogrammed.The patient was pacer dependent and was bradycardic due to the malfunction.The patient would continue to be monitored for trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7558550
MDR Text Key109741815
Report Number2017865-2018-08084
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2016
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000000058
Other Device ID Number05414734508186
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
-
-