• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770501
Device Problems No Display/Image (1183); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Sales rep reported since updating the software on the sr8, the system has been freezing and erasing saved files.
 
Event Description
Sales rep reported since updating the software on the sr8, the system has been freezing and erasing saved files.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of since updating the software on the sr8, the system has been freezing and erasing saved files was unconfirmed.The scanner was evaluated and no issues with the unit freezing or erasing saved files was noted.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.No other functionality issues with the equipment were found during evaluation/servicing.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7563132
MDR Text Key109978875
Report Number3006260740-2018-01199
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/10/2018
Event Location Hospital
Date Manufacturer Received03/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-