ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 05/03/2018
Event Type  malfunction  
Manufacturer Narrative
The customer reported complaint for the platform displayed an error message user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during archive data review and functional testing.Load cells were found defective and a replacement was required.Visual inspection was performed and found no physical damage to the autopulse platform.Following the repair, the autopulse was subjected to the run in test using 95% patient large resuscitation test fixture and passed with no issue or faults observed.Review of the archive data indicated multiple error messages (ua) 07 on the customer reported event date.Historical complaints were reviewed for service information related to the reported complaint and there was one similar complaint reported for autopulse with serial number (b)(4).Ccr (b)(4), reported on (b)(6) 2018.The load cell was replaced to address this complaint.
 
Event Description
During shift check, the autopulse displayed error message user advisory (ua) 07 (discrepancy between load 1 and load 2 too large).The user was unable to clear the issue.No patient was involved with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key7563549
MDR Text Key109888749
Report Number3010617000-2018-00601
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-