• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HL-20 INTEGRATED PERFUSION SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number MCP0.0702625
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).On 2018-05-16: field service technician has been sent for investigation and repair.According to service report# (b)(4) the capasitive level sensor has been replaced.After replacement of the sensor, the unit passed all calibration, functional and safety test performed.The unit was returned to the customer/hospital and cleared for clinical use.Thus the failure could be confirmed.As confirmed by field service technician on 2018-05-29, the sensor was discarded and is therefore no longer available for return for further investigation.Most probable root cause could not be determined exactly.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
As stated by the customer: the level sensor would not function during setup.Error message "sensor fault" occurred.Customer replaced the level sensor with a backup sensor.No patient involvement.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HL-20 INTEGRATED PERFUSION SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7565642
MDR Text Key110180521
Report Number8010762-2018-00199
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCP0.0702625
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-