• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION HOSPITAL CART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION HOSPITAL CART Back to Search Results
Catalog Number 397003-001
Device Problems Failure To Run On AC/DC (1001); Electrical /Electronic Property Problem (1198); Device Emits Odor (1425); Noise, Audible (3273)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because although the hospital cart had a smell, it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion hospital cart will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia authorized distributor, reported that the companion hospital cart made a crackle sound and smell and the docked companion 2 driver was noted to be running on its batteries instead of wall power through the hospital cart.The customer also reported that the patient was subsequently switched to a backup companion 2 driver.There was no reported adverse patient impact.The customer also reported that a system check was performed on the former primary companion 2 driver and it passed all requirements and performed as intended therefore will not be returned.
 
Manufacturer Narrative
Neither the visual or functional investigation could identify a source of the customer-reported crackle sound or the burning smell.Investigation testing determined the root cause of the customer-reported issue of the hospital cart not powering the docked companion 2 driver to be a malfunction of the power supply.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION HOSPITAL CART
Type of Device
HOSPITAL CART
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7566362
MDR Text Key110181841
Report Number3003761017-2018-00210
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Remedial Action Replace
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397003-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age62 YR
-
-