Catalog Number 397003-001 |
Device Problems
Failure To Run On AC/DC (1001); Electrical /Electronic Property Problem (1198); Device Emits Odor (1425); Noise, Audible (3273)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because although the hospital cart had a smell, it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion hospital cart will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia authorized distributor, reported that the companion hospital cart made a crackle sound and smell and the docked companion 2 driver was noted to be running on its batteries instead of wall power through the hospital cart.The customer also reported that the patient was subsequently switched to a backup companion 2 driver.There was no reported adverse patient impact.The customer also reported that a system check was performed on the former primary companion 2 driver and it passed all requirements and performed as intended therefore will not be returned.
|
|
Manufacturer Narrative
|
Neither the visual or functional investigation could identify a source of the customer-reported crackle sound or the burning smell.Investigation testing determined the root cause of the customer-reported issue of the hospital cart not powering the docked companion 2 driver to be a malfunction of the power supply.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|
Search Alerts/Recalls
|