Model Number 1103
Device Problems Positioning Problem (3009); Pumping Problem (3016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/24/2018
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the ventricular assist device (vad) exhibited low flows and associated inflow cannula positioning issue.The vad remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: pump (b)(4) was not returned for evaluation.The reported ¿low flow¿ event was confirmed through log file analysis which revealed multiple low flow alarms that have been logged since (b)(6) 2017 and continuing through (b)(6) 2018.Of note, the site reported an inflow cannula positioning issue, which can be associated with low flows.There is no evidence to suggest that a device malfunction caused or contributed to the reported event.Based on the available information and the risk documentation, the most likely root cause of low flows may be attributed to an inflow cannula positioning issue.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7566628
MDR Text Key110038395
Report Number3007042319-2018-02210
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707000017
UDI-Public00888707000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/30/2019
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-