DEPUY SYNTHES PRODUCTS LLC HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; MOTOR, DRILL, ELECTRIC - HANDPIECE
|
Back to Search Results |
|
Catalog Number EG1A |
Device Problems
Overheating of Device (1437); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/13/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the motor device was making a rattling noise, the bearings within the burr-lock were worn out causing the rattling and overheating.The device also failed the temperature assessment pre-test.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a suplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported that the during pre-surgery, the motor device made a rattling noise.During in-house engineering evaluation, it was determined that the device over heated and the bearings were worn out.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|