DEPUY SYNTHES PRODUCTS LLC HIGH SPEED ELEC G1 HANDPIECE AIR COOLED; MOTOR, DRILL, ELECTRIC - HANDPIECE Back to Search Results
Catalog Number EG1A
Device Problems Overheating of Device (1437); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2018
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the motor device was making a rattling noise, the bearings within the burr-lock were worn out causing the rattling and overheating.The device also failed the temperature assessment pre-test.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a suplemental medwatch will be submitted accordingly.
 
Event Description
It was reported that the during pre-surgery, the motor device made a rattling noise.During in-house engineering evaluation, it was determined that the device over heated and the bearings were worn out.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH SPEED ELEC G1 HANDPIECE AIR COOLED
Type of Device
MOTOR, DRILL, ELECTRIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key7568443
MDR Text Key110035130
Report Number1045834-2018-51055
Device Sequence Number1
Product Code HBC
UDI-Device Identifier845384017134
UDI-Public(01)845384017134(11)140826
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEG1A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Date Manufacturer Received05/18/2018
Date Device Manufactured08/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-