Catalog Number ASKU
Device Problem Low Test Results (2458)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This event occurred in (b)(6).
 
Event Description
The customer stated that they received questionable results for an unspecified number of patient samples tested for ck creatine kinase (ck) and gluc3 glucose hk (glu) on a cobas 6000 c (501) module - c501.The customer stated that the issue occurred when the reagent packs were nearly empty.The customer provided data for one patient sample with an erroneous glu result.The erroneous result was not reported outside of the laboratory.The sample initially resulted with a glu value of < 20 mg/dl.The reagent pack was replaced and after this, the sample was repeated, resulting with a value that was expected.No adverse events were alleged to have occurred with the patient.The c501 analyzer serial number was asked for, but not provided.The issue was resolved after replacement of the reagent pack.The investigation was unable to find a definitive root cause.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLUC3 GLUCOSE HK GEN.3
Type of Device
HEXOKINASE, GLUCOSE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7568545
MDR Text Key110043146
Report Number1823260-2018-01737
Device Sequence Number1
Product Code CFR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K061048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-