Model Number U9525 |
Device Problem
Cut In Material (2454)
|
Patient Problem
No Information (3190)
|
Event Date 10/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Upon triage, the following issue was found: a damaged power cord with exposed copper.
|
|
Manufacturer Narrative
|
One kendall scd express system was triaged and service found the unit to have exposed copper wires on the power cord.The power cord was examined the a cut pattern/mark was observed on the outer shielding and shielding of the copper wire.The potential root cause is customer misuse by running over the cord with hospital bed or by the procedure used to unplug the unit and the procedure of wrapping of the cord around the bed hook.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
|