RESPIRONICS, INC TRILOGY 100; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1054260
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/17/2018
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's blower motor was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key7569319
MDR Text Key110093068
Report Number2518422-2018-01305
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959015364
UDI-Public00606959015364
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Service and Testing Personnel
Type of Report Initial
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054260
Device Catalogue Number1054260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-