Model Number 802100
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 05/11/2018
Event Type  malfunction  
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) observed the ccm to continuously reboot when powered on.It was determined that the flat panel interface node controller (fpinc) printed circuit board (pcb) was the cause of the problem.The product will be sent to service to be brought to manufacturer¿s specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The product surveillance technician (pst) reported that during routine testing of the device at service center, the central control monitor (ccm) kept rebooting.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.It was determined that the single board computer (sbc) printed circuit board assembly (pcba) was the cause of the problem.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7569726
MDR Text Key110181195
Report Number1828100-2018-00293
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number802100
Device Catalogue Number802100
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-