The returned system (of pump, wire, and introducer) was still intact.The guidewire was still inside the pump, and was in a loop within itself.The guidewire and pump were also stuck inside the introducer.The root cause of the pump's inability to cross the aortic valve was unable to be determined.The root cause of the inability to remove the guidewire, along with the pigtail kink, was the loop created by the wire.The root cause of the wire core fracture was most likely the snare used to attempt to remove the wire.There were no other complaints against this lot of pumps or guidewires.
|
An impella 2.5 was placed via the right femoral to support an atherectomy patient in the cath lab.The team lost left ventricle (lv) position and so re-advanced the wire and repositioned the pump within the lv.Upon the team's attempted removal of the guidewire, the pump kinked back upon itself and the wire was unable to be removed.The team therefore attempted to remove the pump and wire in combination, but needed surgical removal techniques to be employed, as contralateral leg snaring of the product was not successful for pump and wire removal.A vessel repair/closure was done to the surgical site and the arteriotomy was closed.
|