• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC INFUSE BONE GRAFT; FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC INFUSE BONE GRAFT; FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET Back to Search Results
Catalog Number 7510400
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Numbness (2415)
Event Type  Injury  
Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Hypoaesthesia it was reported that the patient had underwent unspecified surgery involving rhbmp-2.Post-op, patient reportedly had right lower extremity numbness-intermittent numbness radiating from lateral thigh into foot.Outcome : not recovered/not resolved.
 
Manufacturer Narrative
Additional information: due to imdrf harmonization, some previously submitted device, method, result, and conclusion codes related to this event may have been updated.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Updated ae: worsening bilateral lower extremity pain outcome status: resolved outcome date: (b)(6) 2019 seriousness assessment: non serious sponsor relatedness assessment: the event is unlikely related to devices and surgical procedures.Site relatedness assessment: the event is neither related to devices nor related with surgical procedures.
 
Event Description
Mdr decision corrected to not reportable.No additional supplementals required unless additional information received indicates reportable event.
 
Manufacturer Narrative
Review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunction.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSE BONE GRAFT
Type of Device
FILLER, RECOMBINANT HUMAN BONE MORPHOGENETIC PROTEIN, COLLAGEN SCAFFOLD WITH MET
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer (Section G)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key7571962
MDR Text Key110162424
Report Number1030489-2018-00801
Device Sequence Number1
Product Code NEK
UDI-Device Identifier00681490843805
UDI-Public00681490843805
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/05/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number7510400
Device Lot NumberM111613AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age60 YR
Patient SexFemale
Patient Weight98 KG
-
-