MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1710K; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
Model Number MMT-1710K |
Device Problems
Erratic or Intermittent Display (1182); Device Displays Incorrect Message (2591); Noise, Audible (3273)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/09/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
It was reported that the insulin pump¿s screen flashed and the siren rang.The customer¿s blood glucose level was 100 mg/dl.The insulin pump will be returned for analysis.
|
|
Manufacturer Narrative
|
Insulin pump was received with a constant blank flashing white light on the display and constantly beeping due to vertical cracked lcd controller.Unable to verify missing segments/partial display or perform the functional testings including the self-test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test and displacement test due to insulin pump flashing white light on the display.
|
|