• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 383400
Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reported the enteral feeding pump unexpectedly stopped and will not start back up.There was no patient involvement reported.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for intermittent power.The reported condition was confirmed.A review of the device history record shows that this unit was manufactured in 2014 and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7574516
MDR Text Key110330170
Report Number1282497-2018-00457
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-