Results: the ace68 was partially fractured under the strain relief approximately 2.0 cm from the hub.The ace68 was ovalized approximately 6.0 and 28.0 cm from the hub.Conclusions: evaluation of the returned ace68 revealed a partial fracture under the strain relief near the hub.This damage was likely a result of forcefully manipulating the proximal end of the device at extreme angles.Further evaluation revealed ovalizations on the device proximal shaft.This damage was likely incidental to the reported failure and may have been a result of packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system ace 68 hi-flow kit (kit).During the procedure, while in use with a non-penumbra catheter and a non-penumbra stent retriever device, a ¿pin hole¿ was noticed a little distal to the hub of a penumbra system ace 68 reperfusion catheter (ace68).The physician attempted to tape the ace68 and completed the procedure using the same ace68.There was no report of an adverse effect to the patient.
|