MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125325-12
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Evaluation summary: the device was returned for analysis.The reported hub crack and leak was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.There was no damage or leak noted to the stent delivery system (sds) during the inspection prior to use or during preparation of the device.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely the sds was overly torqued during attachment to the inflation device resulting in the reported hub crack and subsequent hub leak.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the procedure was to treat a 90% stenosed lesion in the mid left anterior descending artery.A 3.25x12mm xience alpine stent delivery system (sds) was prepped without issues.The sds was advanced; however, when the device was being pressurized it was noted that contrast was leaking out of the hub.The pressurization was stopped and all connections were confirmed to be tight.The connections were fine; however, there was a crack at the hub.The sds was removed from the patient anatomy without issues and once outside the anatomy, the balloon appeared as if it had never been inflated.The procedure was successfully completed with a new 3.25x12mm xience alpine sds.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7579942
MDR Text Key110715132
Report Number2024168-2018-04273
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648199424
UDI-Public08717648199424
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2020
Device Catalogue Number1125325-12
Device Lot Number7120761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-