• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 05/07/2018
Event Type  malfunction  
Manufacturer Narrative
The report of the autopulse platform (sn (b)(4)) displaying a system error, out of service, revert to manual cpr message was confirmed during functional testing and archive data review; the root cause was due to a defective processor board.Evaluation of the platform during initial power up, revealed a system error, out of service, revert to manual cpr message on the user control panel.It was indicated that the processor board was defective.The archive data was reviewed and contained a system error 136 (internal parameter corrupted).The autopulse platform is a reusable device and was manufactured on 13 apr 2005.It has exceeded its expected service life of 5 years.The platform is 13 years old and has not seen regular preventive maintenance.As part of routine service during testing, the platform was further examined and found damaged encoder cover.This physical damage found during visual inspection is a characteristic of normal wear and tear for the age of the device.This observation is not related to the reported event.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse platform with serial (b)(4).
 
Event Description
During shift check, the autopulse platform (sn (b)(4)) displayed a "system error, out of service, revert to manual cpr" message upon power up.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7579970
MDR Text Key110459843
Report Number3010617000-2018-00609
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000017
UDI-Public00849111000017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/13/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-