U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 DEPUY/CMW 2G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 DEPUY/CMW 2G; BONE CEMENT : BONE CEMENT Back to Search Results
Model Number 5450-32-500
Device Problems Failure To Adhere Or Bond (1031); Loss of or Failure to Bond (1068)
Patient Problems Inflammation (1932); Not Applicable (3189); No Code Available (3191)
Event Date 05/14/2018
Event Type  Injury  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient was revised to address tibial loosening at the bone to aseptic cement interface.Depuy cement manufacturer.Baseplate well fixed to cement.It was also reported that patient has high bmi and physician believes patient needed longer tibial stem, revised to attune revision rp.Doi: (b)(6) 2016; doi: (b)(6) 2018; left knee.
 
Manufacturer Narrative
Udi: (b)(4).
 
Event Description
Medical records alleges inflammation.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: device history reviewed: no non-conformities on this batch.Final micro and sterility tests passed.Complaints database searched: a complaint database search on the provided lot number found 1 additional report related to implant loosening.Total for lot number: 1 ((b)(4)).Complaints received by cmw in the last 12 months for this issue ¿ by product code: 9.By product family: 19 (8x depuy cmw 1g, 10x depuy cmw 2g, 1x depuy cmw 3g).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.H10 additional narrative: .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPUY/CMW 2G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7581723
MDR Text Key110454197
Report Number1818910-2018-61617
Device Sequence Number1
Product Code MBB
UDI-Device Identifier10603295174271
UDI-Public10603295174271
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 05/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model Number5450-32-500
Device Catalogue Number545032500
Device Lot Number8159148
Was Device Available for Evaluation? No
Date Manufacturer Received07/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ATTUNE MEDIAL DOME PAT 41MM.; ATTUNE PS FEM LT SZ 10 CEM.
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-