• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states that the unit had a blank display.
 
Manufacturer Narrative
The unit was triaged and the reported issue could not be confirmed at this time.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
789 puxing rd
building 10
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN
789 puxing rd
building 10
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7582626
MDR Text Key110575652
Report Number3006451981-2018-00433
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public10884521006614
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/17/2022
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-