Before the capsule endoscopy, the user had confirmed that the recorder had received the capsule endoscopic image of the maj-2027 correctly, and then the patient ingested the subject maj-2027 to start the small intestinal capsule endoscopy.However, the image of the subject maj-2027 was not displayed even after about 30 minutes passed from the start of the capsule endoscopy.The physician retrieved the subject maj-2027 endoscopically and the capsule endoscopy was canceled.Even after retrieve, the image of the subject maj-2027 was not displayed.When the user turned off and on the subject maj-2027 by activator, the image was displayed.The battery was dead when the local staff of olympus received the subject maj-2027 from the user.There was no report of the patient¿s injury other than above.
|
This is a supplemental report for mfr report #8010047-2018-0115 to provide device evaluation results.The subject maj-2027 was returned to olympus medical systems corp.(omsc) for evaluation.Omsc could not confirm the phenomenon.We consider probable causes below.We judge the failure occurred as the device was in other parts except small intestine.This product is for small intestine.The antenna sheet (maj-2031) was not set at correct position.(too high from the correct position).Although the antenna sheet was at correct position when put on, it moved by rolling over or other moving.During endoscope examination, the product was brought close to radio equipment, which caused interference of radio waves.Maj-2027 instruction manual states the corresponding method in case of an abnormality.
|