The customer received a questionable high elecsys ige ii immunoassay result for one patient sample from modular e module serial number (b)(4).The initial result was 46888 iu/ml (>2500 undiluted) and was reported outside of the laboratory.The results for subsequent electrophoresis testing did not agree with this result as no increased immunoglobulin pattern was visible.Therefore, the sample was sent to another laboratory for confirmation.The result by a siemens method was 1002 iu/ml.The customer repeated an aliquot of the same sample on the modular e module and the results were >2500 undiluted and >50000 with an automatic dilution.There was no allegation of an adverse event.From analysis of the calibration and quality control data, a general reagent issue could be excluded.As no sample material was available, the investigation was unable to find a definitive root cause.
|