• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Device Stops Intermittently (1599); Calibration Problem (2890)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the companion 2 driver compressor cycled on and off, the companion 2 driver continued to perform its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the companion 2 driver compressor cycled on and off while supporting a patient.The customer also reported that the patient was switched to a backup companion 2 driver which subsequently had an issue therefore the hospital staff switched the patient back onto the original companion 2 driver.The customer also reported that the patient was extremely stable and continues to be supported by the original companion 2 driver.
 
Manufacturer Narrative
The customer-reported compressor cycling on the companion 2 driver was confirmed via review of the driver's alarm history, and was reproduced during investigation testing.The root cause of the compressor cycling was determined to be a malfunction of the manual pressure regulator, which was unable to maintain the required pressure set point value.Syncardia has a corrective and preventive action (capa) for this issue.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7586804
MDR Text Key111017923
Report Number3003761017-2018-00228
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Remedial Action Replace
Type of Report Initial,Followup
Report Date 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2018
Was the Report Sent to FDA? No
Date Manufacturer Received05/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age52 YR
-
-